The straight halogen lamp and uniquely designed SRA (Secondary Radiation Assist) filter gives shorter measurement time, thanks to fast and uniform heating. The ergonomically designed pan handle eliminates mishaps such as burns when moving a hot sample tray into or out of the unit from either side.

SPECIFICATIONS MS-70 MX-50 MF-50 ML-50
Measurement method 400W straight halogen lamp heating system with SRA-filter and SHS weighing technology
Max. sample weight capacity 71g 51g
Weighing resolution 0.0001g 0.001g 0.002g 0.005g
Moisture content display 0.001%/0.01%/0.1% 0.01%/0.1% 0.05%/0.1%/1% 0.1%/1%
Accuracy over 1g 0.05% 0.10% 0.20% 0.05%
Accuracy over 5g 0.01% 0.02% 0.05% 0.1%
Heating technology Halogen lamp (straight type, 400 Watt, 5000 hours)
Drying temperature 30 — 200°C 50 — 200°C
Measurement programs Standard mode / automatic complete mode / timer mode manual mode / quick m
Units Wet base / dry base / solid content / ratio / weight
Display type Vacuum fluorescent
Sample pan size Ø 85mm